Spray Using the
correct adaptor, spray MK Dent into the drive line. Spray for
approximately 2 seconds. Count 1...... 2...... |

|
|
|
|
Once a week the spindle of the handpiece should be
lubricated. Again, Using the correct adaptor, spray MK Dent
directly into the spindle (where you insert the bur). Spray for
approximately 1 second. Count 1...... |
 |
|
| |
|
| |
|
Run 5 to 10 seconds
Run the handpiece for 5 to 10 seconds or until handpiece achieve
full steady speed to expel excess lubricant.
This should be done at the operatory or at a purge station in the
sterilization room. Count 1...... 2...... 3...... 4..... 5......
6...... 7....... Be sure all excess
lubricant is expelled. Listen to the handpiece. It should run
smoothly at top speed. |

|
|
| |
|
| |
Wipe with alcohol to remove
excess lubricant
Wipe down the exterior of the
handpiece with a
2x2 lightly dampened with alcohol. Use pure
rubbing alcohol only, many over-the-counter "alcohol wipes"
have other solutions added to
them, this can cause spotting of the exterior and severe
damage to the internal parts of the handpiece. |

|
|
| |
|
| |
Bag and Autoclave
Individually bag and sterilize handpiece as
per manufacturer's instructions. Always allow the handpiece to go
through the "Dry Cycle" on your sterilizer. Note: We recommend
using sterilization bags with one side paper and one side plastic |
 |
|
| |
|
| |
|
Let Cool
Remove the handpieces from the
sterilizer and let them cool to room temperature. Running "Hot"
handpieces will severely reduce the life of the turbine.
-Don't Re-Lubricate
-With MK-dent Spray You Lubricate Only Once! |
 |
|
| |
|
| |
Caution:
-
Never wipe or Spray With Surface Disinfectants
-
Never Submerse In Water
-
Never Place Handpiece Into The Ultrasonic Cleaner
|